Máy đo độ dẫn nhiệt Silicone
Model:
HZ-7039
Product
summary:
-
HZ-7039 Silicone thermal
conductivity detector test thin conductor of heat, and solid electrical
insulating materials,thermal grease, resin, rubber, beryllium oxide ceramic,
alumina ceramic materials such as thermal resistance and solid at the interface
thermal contact resistance and thermal conductivity of the material. The test
material for solid-state sheet, the Enclosing box can detect the powder state
materials and paste material.
Standards:
-
MIL-I-49456A (Insulating sheet,
the thermal conductivity of resin, the thermal conductivity of glass fiber
reinforced)
-
GB 5598-85 (Determination of
Beryllium Oxide ceramic thermal conductivity)
-
ASTM D5470-2006 (Thin thermal
conductivity of solid electrical insulating materials, heat transfer
performance testing standards)
-
CPCA4105-2010
Instrument
features:
-
With automatic compression,
automatic thickness measuring device, and connect the computer to achieve
automatic control. The instrument uses temperature gradient detection, improve
the accuracy of the test. Can detect multiple parameters of the thermal
resistance curves under different pressures, optimized mathematical model to
measure the thermal conductivity and thermal resistance of materials and the
interface thermal contact resistance.
Applicable
industry:
-
HZ-7039 Silicone thermal
conductivity detector widely used in universities, research institutes, quality
inspection departments, and production of plant material thermal conductivity
detection.
Technical
parameters:
-
Model: HZ-7039 Silicone thermal
conductivity detector
-
Sample size: ≤Φ30mm
-
Sample thickness: 0.02-20mm
-
Extreme heat temperature range:
RT-299.99℃
-
The extreme cold temperature range:
0-99.00℃
-
Thermal conductivity of the test
range: 0.01~50W/m*k,1~300W/m*k
-
Thermal Resistance Test Range:
0.05~0.000005m2*K/W
-
Pressure measurement range: 0~1000N
-
Displacement measuring range: 0~30.00mm
-
Test accuracy: >3%
-
The sample under vacuum test, to
ensure that the test environment and precision degree of vacuum 0.1MPa.
-
Experimentally
§ Thermal
resistance test specimens of different pressure.
§ Material
thermal conductivity test.
§ The
contact resistance test.
Computer
automatic testing and data printout.