PERFORMANCE

Flow Range
4.00 – 10.00 LPM 
Constant Flow Control
± 5% of set flow at 5-10 LPM
Flow Compensation
Up to 5-hours at:

5 LPM up to 40” H2O back pressure

6 LPM up to 34” H2O back pressure

7LPM up to 24” H2O back pressure

8LPM up to 22” H2O back pressure

9LPM up to 20” H2O back pressure

10LPM up to 14” H2O back pressure
Run Time
8-hours minimum at:

5 LPM up to 26” H2O back pressure

6 LPM up to 23” H2O back pressure

7LPM up to 21” H2O back pressure

8LPM up to 20” H2O back pressure

9LPM up to 15” H2O back pressure

10LPM up to 11” H2O back pressure
Flow Fault
If flow changes exceed 5% within back pressure specifications, fault icon appears. If fault exceeds 30 seconds, pump shuts down. Pump attempts to restart every 3 minutes for up to 30 min.
ENVIRONMENTAL

Temperature Ranges

   Operating 
32°F to 113°F (0°C to 45°C)
   Storage
-4°F to 113°F (-20°C to 45°C)
   Charging
41°F to 104°F (5°C to 40°C)
Humidity Ranges

   Operating
5-85 %RH, non-condensing
   Storage
5-98 %RH, non-condensing
GENERAL

Display
Flow Rate, Total Sample Time,
Total Volume Sampled

Controls
POWER/ENTER, SET/CAL,

BACK/CLEAR, START/RUN/STOP
Indicators
Green LED (“Normal Operation”)
Icons (LCD)
Battery Indicator, Hold, Fault, Set, Prog
Dimensions
3.2W x 5.4H x 2.3D inches

8.2W x 13.7H x 5.8D cm
Weight
20.5 oz. (580 g)
ELECTRICAL

Battery Pack
Removable, Rechargeable 

Nickel-Metal-Hydride
Battery Level Indicator
Icon displays Full, Mid, & Low levels
Interface Connectors
Charging Jack
Charging Time
Under 4 hours
APPROVALS

EN 1232
Type G
Intrinsic Safety

US/Canada NRTL (FM)                         


Class I, Div 1 Groups A, B, C, D;

Class II, Div 1, Groups E, F & G;

Class III, T4
ATEX


 II 1 G

Ex ia IIC T4
CE

Applicable Directives


ATEX 94/9/EC; EMC 2004/108/EC (RFI)

LVD 2006/95/EC