Trang

Thứ Năm, 18 tháng 2, 2016

Máy thử nghiệm phá hỏng do điện áp Model: HZ-7023

Máy thử nghiệm phá hỏng do điện áp
Model: HZ-7023
Product summary:
-          HZ-7023 Voltage breakdown tester mainly applied in solid insulating materials such as: plastics, films, resins, mica, ceramic,glass, insulation paint, and other media in the power frequency voltage or DC voltage breakdown strength and withstanding voltage test of time; The instrument is controlled by computer, can during the test data fast and accurate collection, processing, and can access, display, print.
Equipment security features:
-          Test chamber, the chamber door is open when the power supply of less than a high-voltage transformer input, the high
-          pressure side of the voltage. 100KV test equipment, high voltage electrode from the chamber wall distance greater than
-          370mm, 50KV test equipment, high voltage electrode from the chamber wall distance greater than 250mm, the test even
-          if the contact box wall would not be dangerous.
-          Equipment to install a separate protective earth. Then protect the ground, mainly to reduce sample breakdown of the
-          surrounding  strong electromagnetic interference. You can also avoid control computer out of control.
-           Circuit of the test equipment has a number of protective measures are: over-current protection, voltage protection, leakage
-          protection,short circuit protection, DC test discharge alarm.
-          DC test discharge alarm functions: device equipment to finish the DC test, when you open the pilot door automatic alarm
-          until the alarm after the discharge device discharging device will be automatically canceled. (Note: The DC test does not
-          discharge would be dangerous to personal safety, can not directly pick up electrodes to serve as a reminder to staff discharge
-          so as not to cause injury).
-          The test discharge device for the integration with the host, improving the function of the past alone with a put poles.
Meet the standards:
-          GB1408-2006 insulation materials, electric strength test method, the GB/T1695-2005 vulcanized rubber power frequency voltage breakdown strength and dielectric strength test, the the GB/T3333 power view paper work frequency voltage breakdown test method HG / T 3330 insulated paint film hitwear intensity Determination of GB12656 capacitor paper frequency voltage breakdown test method and ASTM D149 standards.
The main technical requirements:
-          Model: HZ-7024 Rubber vertical France retardant instrument
-          Specimen size: Textile: 80mm×300mm  
-          Burner moving motor power consumption: 10W
-          Ignition control timing: 0 to 999.9 seconds set arbitrarily (using standard 12 seconds). Continued burning time and flame retardant timer: 09999.9 s, accuracy:±0.01s  
-          Dlame height measurement device: Scale high 40mm  
-          Using gas: Propane or butane or liquefied petroleum gas.  
-          Working power: AC220V, 50Hz, 50W   
Weight: about 20kg