Bơm lấy mẫu khí cá nhân: 5
÷  3250 ml/phút
Model: AirChek® 2000
Hãng sản xuất: SKC – Mỹ
Tham khảo tại: http://www.skcinc.com/pumps.asp
| 
   
Flow
  Range 
 | 
  
   
1000
  to 3250 ml/min 
(5 to 500 ml/min requires optional low flow adapter kit)  | 
 
| 
   
Compensation 
Range  | 
  
   
3000
  ml/min at 15 inches water back pressure 
2000 ml/min at 30 inches water back pressure 1000 ml/min at 30 inches water back pressure  | 
 
| 
   
Flow
  Fault 
 | 
  
   
If
  the pump is unable to compensate for longer than 15 seconds due to excessive
  back pressure, the pump enters flow fault mode. During flow fault, the pump
  enters Hold, the fault icon appears on the display during the length of the
  fault, and the accumulated run time display is frozen and retained. After 5
  minutes in flow fault, auto-restart is attempted every 5 minutes up to 10
  times. 
 | 
 
| 
   
Run
  Time 
 | 
  
   
With
  NiMH battery pack, run time is 12+ hours for 2000 ml/min and up to 30 inches
  water back pressure; dependent on sample media used. 
With Battery eliminator, run time is extended.  | 
 
| 
   
Charging
  Time 
(varies with battery capacity and level of discharge)  | 
  
   
6
  to 8.5 hrs with PowerFlex charger 
 | 
 
| 
   
Flow
  accuracy 
 | 
  
   
±
  5% of set-point after calibration to desired flow 
 | 
 
| 
   
Flow
  Control 
 | 
  
   
An
  internal isothermal flow sensor measures flow directly and continuously.
  Sensor readings are used in a flow monitoring algorithm to maintain
  calibrated volumetric flow. In addition, built-in atmospheric temperature and
  pressure sensors provide readings to correct volumetric flow for these
  parameters when they vary from point of calibration. 
 | 
 
| 
   
Atmospheric
  Pressure Accuracy 
 | 
  
   
±
  0.3 in Hg 
 | 
 
| 
   
Timing
  Accuracy 
 | 
  
   
1
  min/mo at 25 C 
 | 
 
| 
   
Operating 
Temperature Range  | 
  
   
32
  to 113 F (0 to 45 C) 
 | 
 
| 
   
Charging 
Temperature Range  | 
  
   
50
  to 113 F (10 to 45 C) 
 | 
 
| 
   
Storage 
Temperature Range  | 
  
   
-4
  to 113 F (-20 to 45 C) 
 | 
 
| 
   
Altitude 
 | 
  
   
The
  pump can apply correction to volumetric flow during sampling for
  weather-related or altitude variations from the atmospheric pressure
  established at calibration up to at least 7500 feet above and 5000 feet below
  sea level. 
 | 
 
| 
   
Battery
  Pack 
 | 
  
   
Removable
  chargeable 4-cell NiMH battery, 4.8 V x 3.5-Ah capacity 
 | 
 
| 
   
Tubing 
 | 
  
   
Requires
  1/4-inch ID tubing 
 | 
 
| 
   
Size 
 | 
  
   
5.6
  x 3 x 2.3 in 
(14.2 x 7.6 x 5.8 cm)  | 
 
| 
   
Weight 
 | 
  
   
22
  oz 
(624 gm)  | 
 
| 
   
Intrinsic
  Safety 
 | 
  
   
UL
  and cUL Listed ATEx and ANZEx models available  
 | 
 
| 
   
Case 
 | 
  
   
Impact
  resistant 
RFI/EMI shielded CE marked  | 
 