Bơm lấy mẫy khí: 10 – 30 LPM
Model:
QuickTake 30
Háng
sản xuất: SKC – Mỹ
Flow
Range
|
10
to 30 L/min
|
||||||||||||||||||||||||||||||||||||||||
Flow
Compensation Accuracy
|
±
5% of set flow
|
||||||||||||||||||||||||||||||||||||||||
Compensating
Flow Back Pressure Range
|
10
L/min at 90 inches water back pressure
20 L/min at 50 inches water back pressure 30 L/min at 15 inches water back pressure |
||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||
Power
|
Rechargeable
lithium-ion (Li-Ion) battery, 7.4-volt, 9.6-Ah capacity, 71.04 Wh, or battery
with AC charger/adapter (100-240 V)
|
||||||||||||||||||||||||||||||||||||||||
Battery
Recharge Time
(varies with battery capacity and level of discharge) |
Approximately
5 hrs with approved charger
|
||||||||||||||||||||||||||||||||||||||||
Typical
Cumulative Run Time‡
|
Spore
Trap cassettes (VersaTrap): 5 hrs at 15 L/min (battery
only)
BioStage viable cascade impactor: 4 hrs at 28.3 L/min (battery only) 25-mm, 1.2-µm MCE filter: 9+ hrs at 10 L/min To achieve, use QuickTake 30 with a fully charged battery and AC charger/adapter. 37-mm, 0.8-µm MCE filter: 14+ hrs at 10 L/min (battery only)
Note: The
AC charger/adapter can be used to extend battery run time, but it is not a
battery eliminator. Therefore, it will not provide indefinite run times.
|
||||||||||||||||||||||||||||||||||||||||
Temperature
|
Operating: 32
to 113 F (0 to 45 C)
Storage: -4 to 95 F (-20 to 35 C) Charging: 32 to 113 F (0 to 45 C) |
||||||||||||||||||||||||||||||||||||||||
Operating
Humidity
|
0
to 95% non-condensing
|
||||||||||||||||||||||||||||||||||||||||
Run
Time Features
|
User-selectable
features, user-adjustable presets. See Operating Instructions.
Preset timed runs: 1, 2, 5, 10, or 15 min Manually set continuous run with manual shut-off: 1 to 999 min (repeats 1 to 999-min runs indefinitely until user stops pump or power supply is depleted) Intermittent sampling: User-set sample periods and sample delays. See Operating Instructions. |
||||||||||||||||||||||||||||||||||||||||
Flow
Fault Features
|
If
the pump is unable to compensate due to excessive back pressure, the pump
will go into flow fault mode. Flow fault features are enabled as the default
setting. Flow fault may be disabled at the user’s option. See
Operating Instructions.
Flow fault: Flow fault displays immediately. Pump shuts off after 5 seconds. Fault restart: Pump attempts restart every 10 seconds up to five times. |
||||||||||||||||||||||||||||||||||||||||
Media
Compatibility
|
BioStage
viable cascade impactors, spore trap cassettes, microvacuum cassettes, and
other samplers that require flows to 30 L/min
|
||||||||||||||||||||||||||||||||||||||||
Tubing
|
Requires
3/8-inch ID tubing
|
||||||||||||||||||||||||||||||||||||||||
Housing
|
ABS
plastic
|
||||||||||||||||||||||||||||||||||||||||
Dimensions
|
9.3
x 8.4 x 3.5 in (23.6 x 21.3 x 8.9 cm)
|
||||||||||||||||||||||||||||||||||||||||
Weight
|
4.8
lbs (2.2 kg)
|
Không có nhận xét nào:
Đăng nhận xét