Thứ Năm, 14 tháng 1, 2016

FLUKE – Máy hiệu chuẩn áp suất Fluke 721-1601 Model: Fluke 721-1601

FLUKE – Máy hiệu chuẩn áp suất Fluke 721-1601
Model: Fluke   721-1601
Hãng sản xuất: FLUKE – MỸ
                                Kết quả hình ảnh cho Fluke   721-1601  FLUKE 
Tính năng Máy hiệu chuẩn áp suất Fluke 721-1601:

The 721 can be configured by selecting either a 16 PSI (1.1 bar) or 36 psi (2.48 bar) low pressure sensor then add any of 7 high pressure ranges including 100, 300, 500, 1000, 1500, 3000 or 5000 psi (6.9, 20, 24.5, 69, 103.4, 200, 345 bar). 
  • Ideal for gas flow calibration (custody transfer) applications
  • Two isolated, stainless steel, pressure sensors with 0.025 % accuracy
  • Pt100 RTD input for temperature measurement, (probe optional)
  • Measures 4 to 20 mA signals
  • Internal 24 V loop power supply can power a transmitter under test
  • Measure up to 30 V dc, check 24 V loop power supplies
  • Extend pressure measurement range with connection to external 700 series pressure modules (29 ranges)
  • Large backlit graphic display can display up to three inputs simultaneously
  • Store five instrument setups for recall and use
Thông số kỹ thuật Máy hiệu chuẩn áp suất Fluke 721-1601:
Specifications 1-Year
Low pressure sensor
Range
Resolution
Accuracy
-14 psi to +16 psi,
-0.97 bar to 1.1 bar
0.001 psi,
0.0001 bar
0.025 % of full scale
High pressure sensor
-12 psi to +100 psi,
-0.83 bar to 6.9 bar
0.01 psi,
0.0001 bar
0.025 % of full scale
General Specifications
Instrument setup recall
Last used on power-up
Environmental
Operating temperature -10 °C to +50 °C
Storage temperature (without batteries) -20 °C to +60 °C
Altitude
2,000 meters
Ingress protection
IP-40
Power requirements
6.0 v dc
Battery life
>35 hours, typical usage
Batteries required
(4) Standard AA cells
Dimensions (HxWxD)
7.9" X 4.3" X 2.3" (20 cm X 11 cm X 5.8 cm)
Weight
03539 kg (1 lb 3 oz)
Connectors/ports pressure
Two, 1/8 in NPT, 700P Series Pressure Module; RTD Probe
Electrical and Temperature Measurement 1-Year
Function
Range
Resolution
Accuracy
mA measure
0 to 24 mA
0.001 mA
±0.015% of rdg
±0.002mA
VDC measure
0 to 30 VDC
0.001V
±0.015% of rdg
±0.002V
*Temperature measure (RTD/ Ohms)
-40 °C to 150 °C
(-40 °F to 302 °F)
0.01 °C, 0.01 °F
±0.015% of rdg
±0.02 ohms
±0.1°C (±0.2°F)
±0.25 degC (±0.45
degF) combined uncertainty when using 720 RTD probe accessory
Loop power supply
24V
N/A
24 mA @ 24V
*Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe available as an accessory
Temperature Effect
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C
Available Pressure Engineering Units
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C, ftH2O@20°C, ftH2O@60°F