FLUKE – Máy hiệu chuẩn áp suất Fluke 719Pro-30G
Model: Fluke 719Pro-30G
Hãng sản xuất: FLUKE – MỸ
Tính năng Máy hiệu chuẩn áp suất Fluke 719Pro-30G:
- Unique integrated electric pump for one handed pressure calibration
 - Test pressure switches easily with easy-to-use switch test function
 - Precision pressure adjust vernier for easy and accurate pressure calibration
 - Ideal accuracy and performance (0.025 %) for reference class transmitter calibration
 - 30 PSI, 150 PSI and 300 PSI pressure ranges
 - Programmable pressure calibrator pump limit:
 - Eliminates over-pressurization
 - Allows the user to program target pressures for pumping
 - Extended pressure measurement range with any of 29 700Pxx Pressure Modules
 - Pump cleanout ports reduce pump failures from moisture and fluid damage
 - Fluid trap included with the unit to mitigate chance of fluid damage in at-risk applications
 - Precision temperature measurement with optional RTD probe
 
Thông số kỹ thuật Máy hiệu chuẩn áp suất Fluke 719Pro-30G:
Specifications 
 | ||||
Pressure Measurement 
 | ||||
Range 
 | 
-12 to 30 PSI /-0.8 to 2 bar 
 | |||
Resolution 
 | 
0.001 PSI, 0.0001 bar 
 | |||
Accuracy (1 year) 
 | 
± 0.025 % of full scale, 6 months ± 0.035 % of full scale,  
1 year Dry Air and non-corrosive gas only  | |||
Temperature effect (all ranges) 
 | 
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C  
No effect on accuracy on all functions from 15 °C to 35 °C  | |||
Function 
 | 
Range 
 | 
Resolution 
 | 
Accuracy (1 year) 
 | 
Comment 
 | 
mA dc (simulation and measurement) 
 | 
0-24 mA dc 
 | 
0.001 mA 
 | 
± 0.015 % of reading ± 2 counts 
 | 
1000 ohm maximum load in mA source, 26 V dc maximum voltage in mA simulate 
 | 
Volts dc (measurement only) 
 | 
30 V dc 
 | 
0.001 V 
 | 
± 0.015 % of reading ± 2 counts 
 | 
No AC voltage measurement, do not exceed 30 V dc 
 | 
Temperature (measurement only, Pt100 RTD) 
 | 
-50 to 150 °C (-50 °F to 302 °F) 
 | 
0.01 °C 
0.01 °F  | 
Temperature ± 0.1 °C (0.2 °F) 
 | 
± 0.25 °C (± 0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory) 
 | 
Loop power supply 
 | 
24 V dc 
 | 
N/A 
 | 
24 mA at 24 V 
 | |
Temperature effect (all functions) 
 | 
Add ± 0.002 % F.S./°C for temps outside of 15 °C to 35 °C 
 | 
No effect on accuracy on all functions from 15 °C to 35 °C 
 | ||
Environmental 
 | ||||
Operating temperature 
 | 
-10 °C to +50 °C 
 | |||
Storage temperature (with batteries) 
 | 
Per battery manufacturer's specification within storage specification below 
 | |||
Storage temperature (without batteries) 
 | 
-20 °C to +60 °C 
 | |||
Power Requirements 
 | ||||
Battery 
 | 
Eight (8) standard Alkaline AA batteries 
 | |||
Battery life 
 | 
300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 PSI; 100 pump cycles to 300 PSI minimum 
 | |||
Physical 
 | ||||
Dimensions (HxWxD) 
 | 
24.13 cm x 11.18 cm x 7 cm (9.5 in x 4.4 in x 2.8 in) 
 | |||
Weight 
 | 
With holster 1.2 kg (2.5 lb) 
 | |||
Connections 
 | ||||
Pressure 
 | 
1/8 inch female NPT 
 | |||
Electrical 
 | 
Standard banana jacks 
 | |||
RTD 
 | 
4 pin miniature circular 
 | |||
External module 
 | 
6 pin miniature circular 
 | |||
Engineering Units (Configurable) 
 | ||||
psi, bar, mbar, kPa, kgcm2, cmH2O@4 °C, cmH2O@20 °C, mH2O@4 °C, mH2O@20 °C, inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, mmHg@0 °C, inHg@0 °C, ftH2O@4 °C, ftH2O@20 °C, ftH2O@60 °F 
 | ||||

Không có nhận xét nào:
Đăng nhận xét