FLUKE – Máy hiệu chuẩn nhiệt độ Fluke 712B
Model: Fluke 712B
Hãng sản xuất: FLUKE – MỸ
Tính năng Máy hiệu chuẩn nhiệt độ Fluke 712B:
- The 712B can measure and simulate (13) different RTD types and resistance
- Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Hanging tool designed in and included with every unit
- Configurable 0% and 100% source settings for quick 25% linearity checks
- Linear ramp and 25% step auto ramp based on 0% and 100% settings
- Dual inputs and backlit display for easy interpretation of measurements
- Power down settings remembered at power up for easy restart of tests
- 1-year and 2-year specifications and traceable certificate of calibration
Thông số kỹ thuật Máy hiệu chuẩn nhiệt độ Fluke 712B:
General Specifications
| |
Maximum voltage applied between any terminal and earth ground or between any two terminals
|
30 V
|
Operating temperature
|
-10 °C to 50 °C
|
Storage temperature
|
- 30 °C to 60 °C
|
Operating altitude
|
2,000 meters
|
Storage altitude
|
12,000 meters
|
Relative humidity (% RH operating without condensation)
|
Non condensing
|
90% (10 °C to 30 °C)
| |
75% (30 °C to 40 °C)
| |
45% (40 °C to 50 °C)
| |
(Without condensation)
| |
Vibration requirements
|
MIL-T-28800E, Class 2
|
Drop text requirements
|
1 meter
|
IP Rating
|
IEC 60529: IP52
|
Electromagnetic environment
|
IEC 61326-1, Portable
|
Safety
|
IEC 61010-1, Max 30 V to earth, Pollution Degree 2
|
Power supply
|
4 AA NEDA 1.5A IEC LR6 batteries
|
Size (H x W x L)
|
52.5 mm x 84 mm x 188.5 mm
|
Weight
|
515 g
|
DC mA Measurement
| |
Resolution
|
0-24 mA
|
Range
|
0.001 mA
|
Accuracy (% of reading + counts)
|
0.010% + 2 μA
|
Temperature coefficient
|
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
|
Ohms Measurement
| |
Ohms range
|
Accuracy (% of reading +counts)
|
0.00 Ω to 400.00 Ω
|
0.015% + 0.05 Ω
|
400.0 Ω to 4000.0 Ω
|
0.015% + 0.5 Ω
|
Note: Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications.
| |
Temperature coefficient
|
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
|
Ohms Source
| ||
Ohms range
|
1.0 Ω to 400.0 Ω
| |
1.00 Ω to 400.00 Ω
| ||
400.0 Ω to 1500.0 Ω
| ||
1500.0 Ω to 4000.0 Ω
| ||
Excitation current from measurement device
|
0.1 mA to 0.5 mA
| |
0.5 mA to 3 mA
| ||
0.05 mA to 0.8 mA
| ||
0.05 mA to 0.4 mA
| ||
Accuracy (% of reading + counts)
|
0.015% + 0.1 Ω
| |
0.015% + 0.05 Ω
| ||
0.015% + 0.5 Ω
| ||
0.015% + 0.5 Ω
| ||
Resolution
|
0.00 Ω to 400.00 Ω
|
0.01 Ω
|
400.0 Ω to 4000.0 Ω
|
0.1 Ω
| |
Temperature coefficient
|
± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
Supports pulsed transmitters and PLCs with pulse times as short as 5 ms |
RTD Input and Output
| |||||||
RTD Type (α)
|
Range (°C)
|
Measure (°C)
|
Source (°C)
| ||||
1-year
|
2-year
|
Source current
|
1-year
|
2-year
| |||
10Ω Pt(385)
|
-200 to 100 °C
|
1.5 °C
|
3 °C
|
1 mA
|
1.5 °C
|
3 °C
| |
100 to 800 °C
|
1.8 °C
|
3.6 °C
|
1 mA
|
1.8 °C
|
3.6 °C
| ||
50Ω Pt(385)
|
-200 to 100 °C
|
0.4 °C
|
0.7 °C
|
1 mA
|
0.4 °C
|
0.7 °C
| |
100 to 800 °C
|
0.5 °C
|
0.8 °C
|
1 mA
|
0.5 °C
|
0.8 °C
| ||
100 Ω Pt(385)
|
-200 to 100 °C
|
0.2 °C
|
0.4 °C
|
1 mA
|
0.2 °C
|
0.4 °C
| |
100 to 800 °C
|
0.015%
+0.18 °C |
0.03%
+0.36 °C |
0.015%
+0.18 °C |
0.03% +0.36 °C
| |||
200 Ω Pt(385)
|
-200 to 100 °C
|
0.2 °C
|
0.4 °C
|
500 μA
|
0.2 °C
|
0.4 °C
| |
100 to 630 °C
|
0.015%
+0.18 °C |
0.03%
+0.36 °C |
0.015%
+0.18 °C |
0.03% +0.36 °C
| |||
500 Ω Pt(385)
|
-200 to 100 °C
|
0.3 °C
|
0.6 °C
|
250 μA
|
0.3 °C
|
0.6 °C
| |
100 to 630 °C
|
0.015%
+0.28 °C |
0.03%
+0.56 °C |
0.015%
+0.28 °C |
0.03% +0.56 °C
| |||
1000 Ω Pt(385)
|
-200 to 100 °C
|
0.2 °C
|
0.4 °C
|
250 μA
|
0.2 °C
|
0.4 °C
| |
100 to 630 °C
|
0.015%
+0.18 °C |
0.03%
+0.36 °C |
0.015%
+0.18 °C |
0.03% +0.36 °C
| |||
100 Ω Pt(3916)
|
-200 to 100 °C
|
0.2 °C
|
0.4 °C
|
1 mA
|
0.2 °C
|
0.4 °C
| |
100 to 630 °C
|
0.015%
+0.18 °C |
0.03%
+0.36 °C |
0.015%
+0.18 °C |
0.03% +0.36 °C
| |||
100 Ω Pt(3926)
|
-200 to 100 °C
|
0.2 °C
|
0.4 °C
|
1 mA
|
0.2 °C
|
0.4 °C
| |
100 to 630 °C
|
0.015 %
+0.18 °C |
0.03%
+0.36 °C |
0.015%
+0.18 °C |
0.03% +0.36 °C
| |||
10 Ω Cu(427)
|
-100 to 260 °C
|
1.5 °C
|
3 °C
|
1 mA
|
1.5 °C
|
3 °C
| |
120 Ω Ni(672)
|
-80 to 260 °C
|
0.15 °C
|
0.3 °C
|
1 mA
|
0.15 °C
|
0.3 °C
| |
50 Ω Cu(427)
|
-180 to 200 °C
|
0.4 °C
|
0.7 °C
|
1 mA
|
0.4 °C
|
0.7 °C
| |
100 Ω Cu(427)
|
-180 to 200 °C
|
0.2 °C
|
0.4 °C
|
1 mA
|
0.2 °C
|
0.4 °C
| |
YSI400
|
15 to 50 °C
|
0.2 °C
|
0.4 °C
|
250 μA
|
0.2 °C
|
0.4 °C
| |
1. Sensor inaccuracies not included.
2. Resolution: 0.1 °C. 3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and Cu50), 0.4 °C (Other RTD types) to the specifications. 4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range). 5. Temperature Coefficient: ±0.05 °C /°C for measure, ± 0.05 °C /°C (<18 °C or >28 °C) for source. 6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms. |
Không có nhận xét nào: