FLUKE - Thiết bị kiểm tra chất lượng nguồn điện Fluke 345 Model: Fluke 1735

FLUKE - Thiết bị kiểm tra chất lượng nguồn điện Fluke 345
Model: Fluke 1735
Hãng sản xuất: FLUKE – MỸ

                                   Kết quả hình ảnh cho Fluke 1735 FLUKE
Tính năng Thiết bị kiểm tra chất lượng nguồn điện Fluke 345:
  • AC/DC current: Clamp-on measurement of ac current up to 1400 A rms and dc current up to 2000 A without breaking the circuit
  • Highest safety rating: 600 V CAT IV power analyzer is rated for use at the service entrance
  • Accurate in noisy environments: Clamp meter performs even with distorted waveforms present on electronic loads with low-pass filter
  • Data logging: Identify intermittent faults by logging any power quality parameter for minutes, or over a month, including harmonics
  • Verify batteries: Direct measurement of dc ripple (%) for battery and dc systems
  • Troubleshoot harmonics: Analyze and log harmonics digitally or graphically
  • Inrush current: Capture and analyze nuisance tripping, from 3 seconds to 300 seconds
  • Easy to use: Easily confirm power meter setup with large backlit color display of waveforms and trends
  • 3-Phase power: Built in capability for balanced loads
  • View graphs and generate reports: Use power analyzer with included Power Log software

Thông số kỹ thuật Thiết bị kiểm tra chất lượng nguồn điện Fluke 345:
98 mm x 52 mm (12 in x 3.75 in x 2 in)

Weight including batteries
 820 g/1.8 lb
Jaw opening
 60 mm
Jaw capacity
 58 mm diameter
Cleaning
 The unit can be cleaned with an Isopropanol impregnated cloth. Do not use abrasives or other solvents.
Electrical data
All accuracies stated at 23 °C ± 1 °C (73.4 °F ± 1.8 °F). See Ambient conditions specifications for temperature coefficients.

Current measurement (dc, dc rms, ac rms)
Measuring range
 0 to 2000 A dc or 1400 ac rms
Autorange facility
 40 A / 400 A / 2000 A
Resolution
 10 mA in 40 A range
 100 mA in 400 A range
 1 A in 2000 A range
Accuracy: DC and dc rms
I > 10 A:
 ± 1.5 % rdg ± 5 digits
I < 10 A:
 ± 0.2 A
Accuracy: AVG
I > 10 A:
 ± 3 % rdg ± 5 digits
I < 10 A:
 ± 0.5 A
Accuracy: Pk
I > 10 A:
 ± 5 % rdg ± 5 digits
I < 10 A:
 ± 0.5 A
Accuracy: AHr
I > 10 A:
 ± 2 % rdg ± 5 digits
I < 10 A:
 ± 0.5 AHr
Accuracy: CF (Crest Factor)
1.1 ≤ CF < 3:
 ± 3 % rdg ± 5 digits
3 ≤ CF < 5:
 ± 5 % rdg ± 5 digits
Resolution:
 0.01
Accuracy: RPL (Ripple)
2 % ≤ RPL < 100 %:
 ± 3 % rdg ± 5 digits
100 % ≤ RPL < 600 %:
 ± 5 % rdg ± 5 digits
Resolution:
 0.1 %
Idc > 5 A, Iac > 2 A

All measurements dc and 15 Hz to 1 kHz
Maximum overload 10,000 A or rms x frequency < 400,000
Amps rms is a true-rms measurement (ac + dc)

Voltage measurement (dc, dc rms, ac rms)
Measuring range
 0 to 825 V dc or ac rms
Autorange facility
 4 V / 40 V / 400 V / 750 V
Resolution
 1 mV in 4 V range
 10 mV in 40 V range
 100 mV in 400 V range
 1 V in 750 V range
Accuracy: DC and dc rms
V > 1 V:
 ± 1 % rdg ± 5 digits
V < 1 V:
 ± 0.02 V
Accuracy: AVG
V > 1 V:
 ± 3 % rdg ± 5 digits
V < 1 V:
 ± 0.03 V
Accuracy: Pk
V > 1 V:
 ± 5 % rdg ± 5 digits
V < 1 V:
 ± 0.03 V
Accuracy: CF (Crest Factor)
1.1 ≤ CF < 3:
 ± 3 % rdg ± 5 digits
3 ≤ CF < 5:
 ± 5 % rdg ± 5 digits
Resolution:
 0.01
Accuracy: RPL (Ripple)
2 % ≤ RPL < 100 %:
 ± 3 % rdg ± 5 digits
100 % ≤ RPL < 600 %:
 ± 5 % rdg ± 5 digits
Resolution:
 0.1 %
Vdc > 0.5 V, Vac > 0.2 V

 All measurements dc and 15 Hz to 1 kHz
Maximum overload 1,000 V rms
Volts rms is a true-rms measurement (ac + dc)
Harmonics
THD (Total Harmonic Distortion)
1 % ≤ THD < 100 %:
 ± 3 % rdg ± 5 digits
100 % ≤ THD < 600 %:
 ± 5 % rdg ± 5 digits
Resolution:
 0.1 %
DF (Distortion Factor)
1 % ≤ DF < 100 %:
 ± 3 % rdg ± 5 digits
Resolution:
 0.1 %
H02 ≤ Vharm < H13:
 ± 5 % ± 2 digits
H13 ≤ Vharm ≤ H30:
 ± 10 % ± 2 digits
 All measurements up to 30th harmonic (40th harmonic for 15 Hz to 22 Hz)
Frequency range of fundamental F0
 15 Hz to 22 Hz and 45 Hz to 65 Hz
Vacrms > 1V
Watts measurement (single- and three-phase) (dc, dc rms, ac rms)
Measuring range
 0 to 1650 kW dc or 1200 kW ac
Autoranging facility
 4 kW, 40 kW, 400 kW, 1650 kW
Resolution
 1 W in 4 kW
 10 W in 40 kW
 100 W in 400 kW
 1 kW in 1200 kW
Accuracy
 2.5 % rdg ± 5 digits
 W1Ø < 2 kW ± 0.08 kW
 W3Ø < 4 kW ± 0.25 kW
VA measurement (single- and three-phase) (dc, dc rms, ac rms)
Measuring range
 0 to 1650 kVA dc or 1200 kVA ac
Autorange facility
 4 kVA, 40 kVA, 400 kVA, 1650 kVA
Resolution
 1 VA in 4 kVA
 10 VA in 40 kVA
 100 VA in 400 kVA
 1 kVA in 1200 kVA
Accuracy
VA > 2 kVA:
 2.5 % rdg ± 5 digits
VA < 2 kVA:
 ± 0.08 kVA
VAR measurement (single- and three-phase)
Measuring range
 0 to 1250 kVAR
Autorange facility
 4 kVAR, 40 kVAR, 400 kVAR, 1200 kVAR
Resolution
 1 VAR in 4 kVAR
 10 VAR in 40 kVAR
 100 VAR in 400 kVAR
 1 kVAR in 1200 kVAR
Accuracy
VAR > 4 kVAR:
 ± 2.5 % rdg ± 5 digits
VAR < 4 kVAR:
 ± 0.25 kVAR
Power factor range
 0.3 < PF < 0.99
Power factor (single- and three-phase)
Measuring range
 0.3 capacitive to 1.0 and 1.0 to 0.3 inductive (72.5° capacitive to 0° and 0° to 72.5° inductive)
Resolution
 0.001
Accuracy
 ± 3°
Frequency range
 15 Hz to 1 kHz
Displacement power factor (single- and three-phase)
Measuring range
 0.3 capacitive to 1.0 and 1.0 to 0.3 inductive (72.5° capacitive to 0° and 0° to 72.5° inductive)
Resolution
 0.001
Accuracy
 ± 3°
Frequency range
 15 Hz to 22 Hz and 45 Hz to 65 Hz
Kilowatt hour (kWHr)
Measuring range
 40,000 kWHr
Autorange facility
 4 kWHr, 40 kWHr, 400 kWHr, 4,000 kWHr, 40,000 kWHr
Resolution
 1 WHr in 4 kWHr Range
 10 WHr in 40 kWHr Range
 100 WHr in 400 kWHr Range
 1 kWHr in 4,000 kWHr Range
 10 kWHr in 40,000 kWHr Range
Accuracy
kWHr > 2 kWHr:
 ± 3 % ± 5 digits
kWHr < 2 kWHr:
 ± 0.08 kWHr
All Watts /VA /VAR /PF measurements
Frequency range:
 DC and 15 Hz to 1 kHz
Current range:
 10 A to 1400 A rms
Voltage range:
 1 V to 825 V rms
Maximum input:
 825 V rms/1400 A rms
Maximum overload:
 1000 V rms/10,000 A All measurements dc and 15 Hz to 1 kHz. Maximum overload 10,000 A or rms x frequency < 400,000
Frequency measurement (from current or voltage sources)
Measuring range
 15 Hz to 1 kHz
Resolution
 0.1 Hz
Accuracy
 15 to 22 Hz ± 0.5 % rdg
 40 Hz to 70 Hz ± 0.5 % rdg
 15 Hz to 1000 Hz ± 1% rdg
Current range
 10 A to 1400 A rms
Voltage range
 1 V to 825 V rms
Scope function
Current measurement
Ranges:
 10 A / 20 A / 40 A / 100 A / 200 A / 400 A / 1000 A / 2000 A
Resolution:
 1 A in 40 A
10 A in 400 A
50 A in 2000 A
Accuracy:
 ± 3 % rdg ± 1 pixel
Maximum overload:
 10,000 A
Voltage measurement
Ranges:
 4 V / 10 V / 20 V / 40 V / 100 V / 200 V / 400 V / 1000 V
Resolution:
 100 mV in 4 V
1 V in 40 V
10 V in 400 V
31.25 V in 1000 V
Accuracy:
 ± 2 % rdg ± 1 pixel
Maximum overload:
 1000 V rms
Frequency range:
 DC and 15 Hz to 600 Hz
Time base
 2.5 ms, 5 ms, 10 ms, 25 ms, 50 ms/div
Refresh rate
 0.5 seconds
Maximum sampling rate
 15.625 kHz
Inrush current function
Ranges
 40 A, 400 A, and 2000 A
Resolution
 10 mA in 40 A range
 100 mA in 400 A range
 1 A in 2000 A range
Accuracy
I > 10 A:
 ± 5 % rdg ± 1 pixel
I < 10 A:
 ± 0.5 A
 All measurements dc and 15 Hz to 1 kHz
Maximum overload
 Maximum overload 10,000 A or rms x frequency < 400,000
 Amps rms is a true-rms measurement (ac + dc)
Capture time
 1 s, 3 s, 10 s, 30 s, 100 s, and 300 s
Maximum sampling rate
 15.625 kHz
Interface
 USB Interface to a PC
 Power Log software for download, analysis and reporting
 345 Upgrade Utility for installing a new firmware version
Logging Memory
Logging areas
 Three areas that can be used individually or combined into one large area
Averaging periods
 1 s, 2 s, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, and custom
Logging Times
Volts and current mode
Average time
 Logging time (1 area)
1 s
 1 h 49 m
2 s
 3 h 38 m
5 s
 9 h 6 m
10 s
 18 h 12 m
30 s
 2 d 6 h 36 m
1 min
 4 d 13 h 12 m
5 min
 22 d 18 h 0 m
10 min
 45 d 12 h 0 m
15 min
 68 d 6 h 0 m
 Logging time (3 areas)
1 s
 5 h 12 m
2 s
 10 h 24 m
5 s
 1 d 2 h 00 m
10 s
 2 d 4 h 00 m
30 s
 6 d 12 h 1 m
1 min
 13 d 0 h 12 m
5 min
 65 d 0 h 15 m
10 min
 130 d 0 h 30 m
15 min
 195 d 0 h 45 m
V & A harmonics mode
Average time
 Logging time (1 area)
1 s
 0 h 34 m
2 s
 1 h 8 m
5 s
 2 h 52 m
10 s
 5 h 44 m
30 s
 17 h 13 m
1 m
 1 d 10 h 26 m
5 m
 7 d 4 h 10 m
10 m
 14 d 8 h 20 m
15 m
 21 d 12 h 30 m
 Logging time (3 areas)
1 s
 1 h 38 m
2 s
 3 h 16 m
5 s
 8 h 11 m
10 s
 16 h 23 m
30 s
 2 d 1 h 11 m
1 m
 4 d 2 h 23 m
5 m
 20 d 11 h 25 m
10 m
 81 d 0 h 50 m
15 m
 121 d 13 h 15 m
single- and three-phase power mode
Average time
 Logging time (1 area)
1 s
 1 h 40 m
2 s
 3 h 21 m
5 s
 8 h 22 m
10 s
 16 h 45 m
30 s
 2 d 2 h 17 m
1 m
 4 d 4 h 35 m
5 m
 20 d 22 h 55 m
10 m
 41 d 21 h 50 m
15 m
 62 d 20 h 45 m
 Logging time (3 areas)
1 s
 4 h 47 m
2 s
 9 h 34 m
5 s
 23 h 57 m
10 s
 1 d 23 h 54 m
30 s
 5 d 23 h 25 m
1 m
 11 d 23 h 25 m
5 m
 59 d 21 h 5 m
10 m
 119 d 18 h 10 m
15 m
 179 d 15 h 15 m