- Insulation test range:
- 1507: 0.01 MΩ to 10 GΩ
- 1503: 0.1 MΩ to 2000 MΩ
- Insulation test voltages:
- 1507: 50 V, 100 V, 250 V, 500 V, 1000 V
- 1503: 500 V, 1000 V
- Save both time and money with Automatic calculation of
Polarization Index and Dielectric Absorption Ratio (only available
on the1507)
- Make repetitive tests simple and easy with the 1507’s Compare
(Pass/Fail) function (only available on the 1507)
- Repetitive or hard-to-reach testing is easy with the remote
test probe
- Live circuit detection prevents insulation test if voltage
> 30 V is detected for added user protection
- Auto-discharge of capacitive voltage for added user protection
- AC/DC voltage: 0.1 V to 600 V
- 200 mA Continuity
- Resistance: 0.01 Ω to 20.00 kΩ
- Save battery power with auto power off
- Read measurements easily with large, backlit display
- CAT IV 600 V overvoltage category rating for added user
protection
- Remote probe, test leads, probes and alligator clips included
with each tester
- Accepts optional Fluke TPAK magnetic hanging system to free
your hands for other work
- Four AA alkaline batteries (NEDA 15 A or IEC LR6) for at least
1000 insulation tests
- One year warranty
|
|
|
|
|
Thông số kỹ thuật Thiết bị đo điện trở cách điện Fluke
1507/1503:
AC/DC Voltage Measurement
|
Accuracy
|
Range
|
600.0 V
|
Resolution
|
0.1 V
|
50 Hz to 400 Hz ± (% of Rdg + Digits)
|
± (2 % + 3)
|
|
Input impedance
|
|
Common mode rejection ratio (1 kΩ
unbalanced):
|
|
> 60 dB at dc, 50 or 60 Hz
|
|
Overload Protection
|
|
Earth bond resistance measurement
|
Range/Resolution
|
20.00 Ω
|
0.01 Ω
|
200.0 Ω
|
0.1 Ω
|
2000 Ω
|
1.0 Ω
|
20.00 kΩ
|
0.01 kΩ
|
|
Accuracy
|
|
Overload Protection
|
|
Open Circuit Test Voltage
|
|
Short circuit current
|
|
Insulation specifications
|
Measurement Range
|
1507:
|
0.01 MΩ to 10 GΩ
|
1503:
|
0.01 MΩ to 2000 MΩ
|
|
Test voltage
|
1507:
|
50 V, 100 V, 250 V, 500 V, 1000 V
|
1503:
|
500 V, 1000V
|
|
Test Voltage Accuracy
|
|
Short circuit current
|
|
Auto Discharge
|
|
Discharge time < 0.5 second for C = 1 µF or less
|
|
Live circuit indicator
|
|
Inhibit test if terminal voltage > 30 V prior to
initialization of test
|
|
Maximum Capacitive Load
|
|
Operable with up to 1 µF load
|
|
Measure Accuracy:
|
50 V:
|
±(3 % + 5)
|
100 V:
|
±(3 % + 5)
|
250 V:
|
±(1.5 % + 5)
|
500 V:
|
±(1.5 % + 5)
|
1000 V:
|
±(1.5 % + 5) to 2000 MΩ,
±(10 % + 3) above 2000 MΩ
|
|
General Specifications
|
Maximum voltage applied to any terminal:
|
|
Storage Temperature
|
|
-40 °C to 60 °C (-40 °F to 140 °F)
|
|
Operating Temperature
|
|
-20 °C to 55 °C (-4 °F to 131 °F)
|
|
Temperature Coefficient
|
|
0.05 x (specified accuracy) per °C for temperatures < 18 °C or
> 28 °C (< 64 °F or > 82 °F)
|
|
Relative Humidity
|
|
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
|
|
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
|
|
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
|
|
Vibration
|
|
Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
|
|
Shock
|
|
1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six
sides, oak floor)
|
|
Electromagnetic compatibility:
|
|
In an RF field of 3 V/M, accuracy = specified accuracy (EN
61326-1:1997)
|
|
Safety
|
|
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO.
61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement category IV 600
V (CAT IV)
|
|
Certifications:
|
|
CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard
IEC/EN 61010-1 2nd Edition
|
|
Batteries
|
|
Four AA batteries (NEDA 15A or IEC LR6)
|
|
Battery Life
|
Insulation test use:
|
Tester can perform at least 1000 insulation tests with fresh
alkaline batteries at room temperature. These are standard tests of 1000 V
into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds
off.
|
Resistance measurements:
|
Tester can perform at least 2500 earth bond resistance
measurements with fresh alkaline batteries at room temperature. These are
standard tests of 1 Ω with a duty cycle of 5 seconds on and
25 seconds off.
|
|
Size
|
|
5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in
L)
|
|
Weight
|
|
IP Rating
|
|
Altitude
|
Operating:
|
2000 m CAT IV 600 V, 3000 m CAT III 600 V
|
Non-operating (storage):
|
12,000 m
|
|
Over-range capability:
|
|
Included accessories:
|
|
TL224 Test Leads, TP74 Test Probes, clips PN 1958654 (red) and PN
1958646 (black), holster and remote probe
|
|